Size:
H x W x D 19.3 x 12.3 x 7.1 inches, 49.0 x 31.2 x 18.0 cm
Weight:
With Power Cartridge 17.4 pounds (7.9 kg)
Without Power Cartridge 13.9 pounds (6.3
kg)
Oxygen
Flow Settings:
Continuous flow settings 0.5 to 3.0 LPM (0.5 liter increments)
Oxygen concentration 91% + 3% @ sea level
Oxygen output pressure 5.0 psig (34.5 kPa) nominal
Pulse flow settings 1.0 to 6.0 (0.5 increments)
Trigger sensitivity Adjustable between –0.25 cm H2O to –0.7
cm H2O
Sound
level:
48 dB(A) at 3.0 LPM continuous flow
40 dB(A) at 3.0 pulse flow setting
Power
specifications:
AC Operation: 100 –
240 VAC 50/60 Hz,
DC Operation: 12 V nominal
Power Cartridge: 195 W-Hrs
Power
Consumption:
48 watts at 1.0 LPM continuous flow
145 watts at 3.0 LPM continuous flow
38 watts at 1.0 pulse flow setting
85 watts at 6.0 pulse flow setting
Nominal
Power Cartridge:
4.0 hours at 1.0 LPM
continuous flow operating time,
1.3 hours at 3.0 LPM
continuous flow,
5.1 hours at 1.0 LPM
pulse flow setting
2.3 hours at 6.0 LPM
pulse flow setting,
Power Cartridge 1.4 to 5.0 hours, dependent upon the flow
set, recharge time setting
Back
up alarm power 9 Volt battery, Operating temperature: 50°
F to 104° F (10° C to 40° C) at up to 28° C dew point, non-condensing,
Altitude operating range 0 – 13,123 feet (0 – 4000 meters)
Alarms:
Loss of power, Low Power Cartridge, Low Therapeutic Oxygen
Output, Oxygen flow outside normal limits, No inspiration
detected in Pulse Flow, Unit Malfunction.
Note:
For your safety, the Eclips is equiped with Oxygen Monitor
(OCI).
|